Solution to liquid chromatography flat peak failure

Fault phenomenon:

Pingding Peak

possible reason:

(1) Column overload

(2) Sample, mobile phase problem

(3) Pollution detection of the optical pool and its lens, pool window and other optical accessories

(4) Detector signal acquisition failure

Method of exclusion:

(1) Reduce the injection volume

(2) Check the sample and mobile phase, and take the new sample and mobile phase replacement method to check

(3) Online recoil flow pool (do not handle optical components without authorization, please contact RIGOL engineer if you have any questions)

(4) Please contact Shanghai 屹利 engineer for help

Shanghai Jiapeng Technology Co., Ltd. specializes in: UV analyzer, three-use UV analyzer, dark box UV analyzer, black box UV analyzer, black box UV analyzer, portable UV analyzer, three-use UV analyzer, dark box , UV detectors, partial collectors, constant current pumps, peristaltic pumps, gel imaging systems, gel imaging analysis systems, chemiluminescence imaging analysis systems, photochemical reactors, vortex mixers, vortex mixers, glass chromatography columns, Gradient mixer, gradient mixer, nucleic acid protein detector, glass chromatography column, fluorescent whitening agent analyzer, fraction collector, rubber cutting instrument, blue gel cutting instrument, chromatography system and other products. Welcome to inquire.