Analyze the maintenance and precautions of the closed flash point tester

Analyze the maintenance and precautions of the closed flash point tester

Analyze the maintenance and precautions of the closed flash point tester

1. Keep the instrument clean to prevent acid, alkali, oil, moisture and other corrosion.

2. After the experiment is completed, turn the adjustment knob counterclockwise to the end point, turn off the power, and pull out the power plug.

3. During the test, the cover of the instrument is partially hot, and contact should be avoided to facilitate the safe use of the operator and the instrument.

4. When performing step B, replace the 105 rpm motor with a 250 rpm motor. When replacing, please tighten the relevant screws.

5. Before the ignition, please pay attention to tighten the valve to avoid excessive flame during ignition. However, due to the slow moving speed of the gas in the gas pipeline, the time may be longer, please be patient.

6. When the instrument fails, please ask the professional to repair it. Do not disassemble it or contact us.

7, common faults and causes

malfunction
the reason
The power indicator is off.
Heater off power switch is broken
Adjust knob voltmeter pointer does not move
The heater is not hot
Heater voltmeter is bad
Solid state regulator is bad
Heater is not adjustable
Solid state voltage regulator breakdown