Instructions for Elisa Kit for Rat Insulin (INS)
This kit is for research use only.
Detection range: 96T
0 mU / L -20 mU / L
purpose of usage:
This kit is used to determine the content of insulin (INS) in rat serum samples.
Experimental principle
Rat insulin (INS) Elisa kit instructions This kit uses the double antibody sandwich method to determine the level of rat insulin (INS) in the specimen. With purified rat insulin
(INS) antibody is coated on the microwell plate to make solid phase antibody. Insulin (INS) is added to the monoclonal antibody-coated microwells in turn, followed by HRP
The labeled insulin (INS) antibody binds to form an antibody-antigen-enzyme-labeled antibody complex. After thorough washing, the substrate TMB is added
Color rendering. TMB is converted into blue under the catalysis of HRP enzyme, and into the final yellow under the action of acid. Deep color
There is a positive correlation between the shallow and the insulin (INS) in the sample. Measure the absorbance (OD value) at 450nm with a microplate reader, pass
The standard curve calculates the concentration of rat insulin (INS) in the sample.
Kit composition
1 30 times concentrated washing solution 20ml × 1 bottle 7 stop solution 6ml × 1 bottle
2 Enzyme label reagent 6ml × 1 bottle 8 standard (40 mU / L) 0.5ml × 1 bottle
3 Enzyme label coated plate 12 wells × 8 strips 9 standard dilutions 1.5ml × 1 bottle
4 Sample diluent 6ml × 1 bottle 10 instructions 1 copy
5 Developer A solution 6ml × 1 bottle 11 2 sealing film
6 Developer B liquid 6ml × 1 / bottle 12 1 sealed bag
Specimen requirements
1. Specimens are extracted as soon as possible after collection, and extraction is performed according to relevant literature, and experiments should be conducted as soon as possible after extraction. If not
Perform the test immediately. The specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided
2. The samples containing NaN3 cannot be detected because NaN3 inhibits horseradish peroxidase (HRP) activity.
Steps
1. Dilution of standard products: this kit provides one original standard product, the user can dilute in a small test tube according to the following chart
release.
20 mU / L No. 5 standard 150μl original standard added 150μl standard dilution
10 mU / L No. 4 standard 150 μl No. 5 standard added 150 μl standard dilution
5 mU / L No. 3 standard 150 μl No. 4 standard added 150 μl standard diluent
2.5 mU / L No. 2 standard 150μl No. 3 standard added 150μl standard diluent
1.25 mU / L No. 1 standard 150μl No. 2 standard added 150μl standard diluent
2. Add sample: set up blank wells (the blank control wells do not add samples and enzyme reagents, the rest of the operation is the same), standard well
Sample hole to be tested. Add 50μl of the standard on the enzyme-coated plate accurately, and add 40μl of sample diluent to the sample well.
Then add 10μl of the sample to be tested (the final dilution of the sample is 5 times). Add the sample to the bottom of the well of the microtiter plate
Do not touch the wall of the hole, shake gently to mix.
3. Incubation: seal the plate with the sealing film and incubate at 37 ° C for 30 minutes.
4. Mixing solution: Dilute 30 times concentrated washing liquid with distilled water 30 times and reserve
5. Washing: Carefully peel off the sealing film, discard the liquid, spin dry, fill each well with washing liquid, let stand for 30 seconds and then discard, so
Repeat 5 times and pat dry.
6. Add enzyme: add 50μl of enzyme label reagent to each well, except blank well.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: add 50μl of developer A to each well, and then add 50μl of developer B, mix gently, and develop color at 37 ℃ in the dark
15 minutes.
10. Termination: Add 50μl of stop solution to each well to terminate the reaction (at this time, the blue color turns to yellow).
11. Determination: Measure the absorbance (OD value) of each well in sequence with the blank air conditioner at zero and 450nm wavelength. Determination should be terminated
Within 15 minutes after the solution.
Summary of operating procedures:
Calculation
Taking the concentration of the standard as the abscissa and the OD value as the ordinate, draw a standard curve on the coordinate paper, according to the sample
The corresponding concentration of OD value is found by the standard curve; multiplied by the dilution factor; or the standard concentration and OD value are used to calculate the standard
The linear regression equation of the quasi-curve, the OD value of the sample is substituted into the equation, the sample concentration is calculated, and then multiplied by the dilution factor,
This is the actual concentration of the sample.
Precautions
1. The kit should be taken out of the refrigerated environment and equilibrated at room temperature for 15-30 minutes before it can be used.
After use, the slats should be stored in sealed bags.
2. Crystals may be precipitated in the concentrated washing liquid, which can be heated and dissolved in the water bath during dilution, and the results will not be affected during washing.
3. The sample adder should be used in each step of sample addition, and its accuracy should be regularly checked to avoid test errors. One sample loading time is best
Control within 5 minutes, if the number of specimens is large, it is recommended to use a volley gun to add samples.
4. Please make the standard curve at the same time of each measurement, it is better to make the complex hole. If the content of the test substance in the specimen is too high (sample OD value
Is greater than the OD value of the first well of the standard product), please dilute it with a certain multiple (n times) of the sample diluent before measuring.
When calculating, please multiply the total dilution factor (× n × 5).
5. The sealing film is limited to one-time use to avoid cross contamination.
6. Please keep the substrate away from light.
7. Strictly follow the instructions, and the test results must be determined based on the reading of the microplate reader.
8. All samples, washing liquid and various wastes should be treated as infectious agents.
9. The components of different batches of this reagent shall not be mixed.
Storage conditions and validity period
1. Kit storage :; 2-8 ℃.
2. Validity: 6 months Rat insulin (INS) Elisa kit instructions
Easy operation Food Printer with edible ink was manufactured as a homemade printer with economic print head, food printer is customized for entry-level customers and coffee shops. Easy operation, high printing speed and low cost will attract more clients to come. It will creat more value for you.
Name
Food printer
Print head
DX5
Print size
21*38cm
Print color
CMYK+LC LM
Print speed
A4 photo/163s
Printing resolution
5760*1440dpi
Nozzles
90*6=540
Interface
USB2.0
Net weight/Gross weight
35Kg/42Kg
Printer size
57*48*44cm
Height adjustment
Automatic
Working power
110-220V 50-60Hz 35-75W
Operation system
Windows 10/8/7/2000/XP/VISTA eta
Temperature
5-35º
Food Printer
Food Printer,Food Printing Machine,Digital Food Printer,Automatic Food Printer
Shenzhen Refinecolor Technology Co., LTD. , https://www.rfcprinter.com